Techniques
Common Elements
Certain areas on various set-up pages are common to most techniques they are:
Each
set-up page has a diagram of any cell connections required for the test to function
correctly
optional connections are shown in light grey.
All potentials measured are displayed with respect to the reference electrode.
This
section allows the precise control of internal count resistors at the beginning and during
a test. The Auto ranging during test if checked allows the instrument to change to
the best count resistor whilst the test is running. Also at the start of the test the
internal count resistor can be set to a certain resistor
or to be automatically selected
before the test is started. These two options can be combined in any order
for example it
is possible to automatically find the best count resistor at the start of the test
but to
fix on this resistor during the test (Auto ranging during test off
and At Start
to Auto).
![]()